OtO Photonics' SilverBullet (SB) series are a new generation of ultra-compact Czerny-Turner cavity spectrometers featuring a single concave mirror in place of separate collimator and focussing mirror components.
This innovative arrangement provides the sensitivity and resolution benefits of a Czerny-Turner cavity design in an extremely compact form-factor with an overall size (including connectors and control boards) of only 40mm x 36.3mm x 25.1mm (~36cm³ ).
Despite its compact form-factor SilverBullet features an extremely rigid optical bench ensuring a platform with excellent thermal, humidity, shock and vibration stability.
The high speed 1024-pixel CMOS sensor and 8-bit 8051 microcontrollers provide high optical resolution and sensitivity along with short integration time, fast readout speed, high accuracy trigger timing and low power consumption.
Description | SB3134 SB4134 |
Sensor | Hamamatsu S13014 1024 Pixel High Speed CMOS |
Dark Noise | 19 (Max.) |
Dynamic Range | 4000:1 |
SNR | 350 |
Integration Time | 17µs to 24s |
Optical Parameters | f/#: 4.5 NA: 0.11 |
Dimensions | 40 x 36.3 x 25.1 mm |
SilverBullet can be specified with a wide choice of grating and spectral range along with two slit widths, allowing users to select the appropriate wavelength range, resolution and sensitivity.
Model Name |
Groove Density (g/mm) |
Blaze Angle (Optimum Efficiency Wavelength) |
Spectral Range (nm) |
Resolution (nm, FWHM) |
||
25µm | 50µm | 100µm | ||||
VNIR7B | 300 | 500nm | 300-1100 | <6.5 | <10.8 | <21.5 |
DUVN5 | 400 | 550nm | 250-1070 | <5.0 | <7.8 | ~ |
FUV16 | 500 | 240nm | 180-800 | <4.2 | <6.2 | <11.4 |
DUV2A | 500 | 240nm | 200-850 | <4.2 | <6.2 | <11.4 |
FUVA | 500 | 330nm | 180-850 | <4.2 | <6.2 | <11.4 |
DUV2B | 500 | 330nm | 200-850 | <4.2 | <6.2 | <11.4 |
V4E | 500 | 330nm | 300-850 | <4.2 | <6.2 | <11.4 |
500 | 560nm | 180-850 | <4.2 | <6.2 | <11.4 | |
V28 | 500 | 560nm | 380-800 | <4.2 | <6.2 | <11.4 |
FUV19 | 600 | 250nm | 180-700 | <3.5 | <5.8 | ~ |
DUV2 | 600 | 250nm | 200-850 | <3.5 | <5.8 | ~ |
700 | 530nm | 180-600 | <3.0 | <5.0 | ~ | |
768 | 420nm | 180-550 | <2.7 | <4.5 | ~ | |
FUV15 | 900 | 200nm | 180-470 | <2.3 | <3.9 | ~ |
1000 | 250nm | 180-430 | <2.1 | <3.5 | ~ | |
1200 | 200nm | 180-350 | <1.7 | <2.9 | ~ |
Over time we expect additional grating and slit width options to become available with application-specific spectral range, sensitivity and resolution characteristics. Please feel free to discuss your application's needs to see if we can match to existing configurations or develop an optimised solution.
![]() |
SilverBullet product brief |
![]() |
SilverBullet data sheet |
Power and communication with SB-Series spectrometers is via high speed micro-USB with the addition of 6 user-programmable digital I/Os.
The inclusion of OtO's SpectraSmart software ensures that SilverBullet users can be up-and-running within minutes whilst the availability of a Software Development Kit for Windows and example code for Linux facilitates integration into a broad range of OEM instruments.
For more information on SilverBullet and to discuss your application, please Contact Us.